Ma, D;
Mason, C;
Ghoreishizadeh, SS;
(2018)
A wireless system for continuous in-mouth pH monitoring.
In:
Proceedings of the IEEE Biomedical Circuits and Systems Conference (BioCAS) 2017.
IEEE: Turin, Italy.
Preview |
Text
Ma_IEEE_BIOCAS_2017.pdf - Accepted Version Download (2MB) | Preview |
Abstract
A first prototype of a Smart Orthodontic Bracket (SOB) for continuous in-mouth pH monitoring is presented in this paper. The SOB system uses an Iridium Oxide (IrOx) pH sensor with 68.8 mV/pH measured sensitivity. It is powered through Near Field Communications (NFC) using a smartphone from a distance of 3.5 cm. The system resolves pH variations as small as 0.15 within a wide range of pH. The system is encapsulated in bio-compatible epoxy resin and successfully used to measure the pH in saliva.
Type: | Proceedings paper |
---|---|
Title: | A wireless system for continuous in-mouth pH monitoring |
Event: | 2017 IEEE Biomedical Circuits and Systems Conference (BioCAS) |
ISBN-13: | 9781509058037 |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1109/BIOCAS.2017.8325556 |
Publisher version: | https://doi.org/10.1109/BIOCAS.2017.8325556 |
Language: | English |
Additional information: | This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions. |
Keywords: | Electrodes, Antenna measurements, Integrated circuits, Wireless communication, Sensors, Mouth, Wires |
UCL classification: | UCL UCL > Provost and Vice Provost Offices UCL > Provost and Vice Provost Offices > UCL BEAMS UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Electronic and Electrical Eng |