UCL Discovery Stage
UCL home » Library Services » Electronic resources » UCL Discovery Stage

Noninvasive Instrument-based Tests for Detecting and Measuring Vitreous Inflammation in Uveitis: A Systematic Review

Liu, X; Hui, BT; Way, C; Beese, S; Adriano, A; Keane, PA; Moore, DJ; (2020) Noninvasive Instrument-based Tests for Detecting and Measuring Vitreous Inflammation in Uveitis: A Systematic Review. Ocular Immunology and Inflammation 10.1080/09273948.2020.1799038. (In press). Green open access

[thumbnail of Noninvasive Instrument based Tests for Detecting and Measuring Vitreous Inflammation in Uveitis A Systematic Review.pdf]
Preview
Text
Noninvasive Instrument based Tests for Detecting and Measuring Vitreous Inflammation in Uveitis A Systematic Review.pdf - Published Version

Download (1MB) | Preview

Abstract

PURPOSE: This systematic review aims to identify instrument-based tests for quantifying vitreous inflammation in uveitis, report the test reliability and the level of correlation with clinician grading. METHODS: Studies describing instrument-based tests for detecting vitreous inflammation were identified by searching bibliographic databases and trials registers. Test reliability measures and level of correlation with clinician vitreous haze grading are extracted. RESULTS: Twelve studies describing ultrasound, optical coherence tomography (OCT), and retinal photography for detecting vitreous inflammation were included: Ultrasound was used for detection of disease features, whereas OCT and retinal photography provided quantifiable measurements. Correlation with clinician grading for OCT was 0.53-0.60 (three studies) and for retinal photography was 0.51 (1 study). Both instruments showed high inter- and intra-observer reliability (>0.70 intraclass correlation and Cohen's kappa), where reported in four studies. CONCLUSION: Retinal photography and OCT are able to detect and measure vitreous inflammation. Both techniques are reliable, automatable, and warrant further evaluation.

Type: Article
Title: Noninvasive Instrument-based Tests for Detecting and Measuring Vitreous Inflammation in Uveitis: A Systematic Review
Location: England
Open access status: An open access version is available from UCL Discovery
DOI: 10.1080/09273948.2020.1799038
Publisher version: http://dx.doi.org/10.1080/09273948.2020.1799038
Language: English
Keywords: Systematic review, diagnostic test, imaging, optical coherence tomography, retinal photography, ultrasound, uveitis, vitreous, vitreous inflammation, vitritis
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Brain Sciences > Institute of Ophthalmology
URI: https://discovery-pp.ucl.ac.uk/id/eprint/10112834
Downloads since deposit
1,620Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months