Norrish, G; Chubb, H; Field, E; McLeod, K; Ilina, M; Spentzou, G; Till, J; ... Kaski, JP; + view all (2021) Clinical outcomes and programming strategies of implantable cardioverter-defibrillator devices in paediatric hypertrophic cardiomyopathy: a UK National Cohort Study. EP Europace , 23 (3) pp. 400-408. 10.1093/europace/euaa307. Green open access

[thumbnail of Norrish_ICD manuscript revision clean to submit  2.pdf]
Preview
Text
Norrish_ICD manuscript revision clean to submit 2.pdf - Accepted Version

Download (719kB) | Preview

Abstract

AIMS: Sudden cardiac death (SCD) is the most common mode of death in paediatric hypertrophic cardiomyopathy (HCM). This study describes the implant and programming strategies with clinical outcomes following implantable cardioverter-defibrillator (ICD) insertion in a well-characterized national paediatric HCM cohort. METHODS AND RESULTS: Data from 90 patients undergoing ICD insertion at a median age 13 (±3.5) for primary (n = 67, 74%) or secondary prevention (n = 23, 26%) were collected from a retrospective, longitudinal multi-centre cohort of children (<16 years) with HCM from the UK. Seventy-six (84%) had an endovascular system [14 (18%) dual coil], 3 (3%) epicardial, and 11 (12%) subcutaneous system. Defibrillation threshold (DFT) testing was performed at implant in 68 (76%). Inadequate DFT in four led to implant adjustment in three patients. Over a median follow-up of 54 months (interquartile range 28-111), 25 (28%) patients had 53 appropriate therapies [ICD shock n = 45, anti-tachycardia pacing (ATP) n = 8], incidence rate 4.7 per 100 patient years (95% CI 2.9-7.6). Eight inappropriate therapies occurred in 7 (8%) patients (ICD shock n = 4, ATP n = 4), incidence rate 1.1/100 patient years (95% CI 0.4-2.5). Three patients (3%) died following arrhythmic events, despite a functioning device. Other device complications were seen in 28 patients (31%), including lead-related complications (n = 15) and infection (n = 10). No clinical, device, or programming characteristics predicted time to inappropriate therapy or lead complication. CONCLUSION: In a large national cohort of paediatric HCM patients with an ICD, device and programming strategies varied widely. No particular strategy was associated with inappropriate therapies, missed/delayed therapies, or lead complications.

Type: Article
Title: Clinical outcomes and programming strategies of implantable cardioverter-defibrillator devices in paediatric hypertrophic cardiomyopathy: a UK National Cohort Study
Location: England
Open access status: An open access version is available from UCL Discovery
DOI: 10.1093/europace/euaa307
Publisher version: https://doi.org/10.1093/europace/euaa307
Language: English
Additional information: This version is the author accepted manuscript. For information on re-use, please refer to the publisher’s terms and conditions.
Keywords: Childhood, Hypertrophic cardiomyopathy, Implantable cardioverter-defibrillator, Programming
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science
URI: https://discovery-pp.ucl.ac.uk/id/eprint/10116284
Downloads since deposit
5,548Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item