Abstract
Background: The subcutaneous implantable cardioverter-defibrillator (S-ICD) is an alternative to conventional transvenous ICD (TV-ICD) therapy to reduce lead complications. Objective: To evaluate outcomes in channelopathy vs patients with structural heart disease in the EFFORTLESS-SICD Registry and with a previously reported TV-ICD meta-analysis in channnelopathies. Methods: The EFFORTLESS registry includes 199 patients with channelopathies (Brugada syndrome 83, long QT syndrome 24, idiopathic ventricular fibrillation 78, others 14) and 786 patients with structural heart disease. Results: Channelopathy patients were younger (39 ± 14 years vs 51 ± 17 years; P < .001) with left ventricular ejection fraction 59% ± 9% vs 41% ± 18% (P < .001). The complication rate (follow-up: 3.2 ± 1.5 years vs 3.0 ± 1.5 years) was similar: 13.6% vs 11.2% (P = .42). Appropriate shocks rates were 9.5% vs 10.8% (P = .70), with shocks for monomorphic ventricular tachycardia being 2.0% vs 6.9% (P < .02) and for polymorphic ventricular tachycardia/ventricular fibrillation (VT/VF) 8.0% vs 5.7% (P = .30). Conversion effectiveness of VT/VF episodes was similar: 36 of 37 (97.3%) vs 151 of 155 (97.4%, P = .59). VT/VF storm event (2% vs 0.9%, P = .33) and lower inappropriate shock (IAS) (8.5% vs 12.5%, P = .12) rates were statistically similar between channelopathy and non-channelopathy patients, with 45.5% channelopathy vs 31.4% non-channelopathy patients managed with a conditional zone > 200 beats per minute (P = .0002). Annualized appropriate shock, IAS, and complication rates appear to be lower for the S-ICD vs meta-analysis TV-ICD patients, particularly lead complications. Conclusion: EFFORTLESS demonstrates similar S-ICD efficacy and a nonsignificant, lower rate of IAS in channelopathy patients as compared to structural heart disease. Comparable IAS rates were achieved with the device programmed to higher rates for channelopathy patients.
Type: | Article |
---|---|
Title: | Evaluation of subcutaneous implantable cardioverter-defibrillator performance in patients with ion channelopathies from the EFFORTLESS cohort and comparison with a meta-analysis of transvenous ICD outcomes |
Location: | United States |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1016/j.hroo.2020.10.002 |
Publisher version: | http://dx.doi.org/10.1016/j.hroo.2020.10.002 |
Language: | English |
Additional information: | Copyright © 2020 Published by Elsevier Inc. on behalf of Heart Rhythm Society. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/ licenses/by-nc-nd/4.0/). |
Keywords: | Arrhythmia, Channelopathy, Implantable cardioverter-defibrillator, Subcutaneous ICD, Sudden cardiac death, Ventricular arrhythmias |
UCL classification: | UCL UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science > Clinical Science |
URI: | https://discovery-pp.ucl.ac.uk/id/eprint/10132001 |