(2022)
The effect of a variable focal spot size on the contrast channels retrieved in edge illumination x-ray phase contrast imaging.
Scientific Reports
, 12
, Article 3354. 10.1038/s41598-022-07376-0.
Preview |
Text
s41598-022-07376-0.pdf Download (3MB) | Preview |
Abstract
Multi-modal X-ray imaging allows the extraction of phase and dark-field (or “Ultra-small Angle Scatter”) images alongside conventional attenuation ones. Recently, scan-based systems using conventional sources that can simultaneously output the above three images on relatively large-size objects have been developed by various groups. One limitation is the need for some degree of spatial coherence, achieved either through the use of microfocal sources, or by placing an absorption grating in front of an extended source. Both these solutions limit the amount of flux available for imaging, with the latter also leading to a more complex setup with additional alignment requirements. Edge-illumination partly overcomes this as it was proven to work with focal spots of up to 100 micron. While high-flux, 100 micron focal spot sources do exist, their comparatively large footprint and high cost can be obstacles to widespread translation. A simple solution consists in placing a single slit in front of a large focal spot source. We used a tunable slit to study the system performance at various effective focal spot sizes, by extracting transmission, phase and dark-field images of the same specimens for a range of slit widths. We show that consistent, repeatable results are obtained for varying X-ray statistics and effective focal spot sizes. As the slit width is increased, the expected reduction in the raw differential phase peaks is observed, compensated for in the retrieval process by a broadened sensitivity function. This leads to the same values being correctly retrieved, but with a slightly larger error bar i.e. a reduction in phase sensitivity. Concurrently, a slight increase in the dark-field signal is also observed.
Type: | Article |
---|---|
Title: | The effect of a variable focal spot size on the contrast channels retrieved in edge illumination x-ray phase contrast imaging |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1038/s41598-022-07376-0 |
Publisher version: | http://www.nature.com/srep/index.html |
Language: | English |
Additional information: | Open Access: This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. |
UCL classification: | UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng UCL > Provost and Vice Provost Offices > UCL BEAMS UCL |
URI: | https://discovery-pp.ucl.ac.uk/id/eprint/10142831 |
Archive Staff Only
![]() |
View Item |