UCL Discovery Stage
UCL home » Library Services » Electronic resources » UCL Discovery Stage

Bend-insensitive Fiber Optic Ultrasonic Tracking Probe for Cardiovascular Interventions

Mathews, Sunish; Little, Callum; Zhang, Edward; Beard, Paul; Mastracci, Tara; Rakhit, Roby; Desjardins, Adrien; (2023) Bend-insensitive Fiber Optic Ultrasonic Tracking Probe for Cardiovascular Interventions. Medical Physics 10.1002/mp.16334. (In press). Green open access

[thumbnail of Published version]
Preview
Text (Published version)
Medical Physics - 2023 - Mathews - Bend‐insensitive fiber optic ultrasonic tracking probe for cardiovascular interventions.pdf

Download (3MB) | Preview
[thumbnail of Supplementary Material]
Preview
Text (Supplementary Material)
Med_Phys_Supplementary_Materials-Revised_Clean_version.pdf - Supplemental Material

Download (797kB) | Preview

Abstract

Background: Transesophageal echocardiography (TEE) is widely used to guide medical device placement in minimally invasive cardiovascular procedures. However, visualization of the device tip with TEE can be challenging. Ultrasonic tracking, enabled by an integrated fiber optic ultrasound sensor (FOUS) that receives transmissions from the TEE probe, is very well suited to improving device localization in this context. The problem addressed in this study is that tight deflections of devices such as a steerable guide catheter can result in bending of the FOUS beyond its specifications and a corresponding loss of ultrasound sensitivity. Purpose: A bend-insensitive FOUS was developed, and its utility with ultrasonic tracking of a steerable tip during TEE-based image guidance was demonstrated. Methods: Fiberoptic ultrasound sensors were fabricated using both standard and bend insensitive single mode fibers and subjected to static bending at the distal end. The interference transfer function and ultrasound sensitivities were compared for both types of FOUS. The bend-insensitive FOUS was integrated within a steerable guide catheter, which served as an exemplar device; the signal-to-noise ratio (SNR) of tracking signals from the catheter tip with a straight and a fully deflected distal end were measured in a cardiac ultrasound phantom for over 100 frames. Results: With tight bending at the distal end (bend radius < 10 mm), the standard FOUS experienced a complete loss of US sensitivity due to high attenuation in the fiber, whereas the bend-insensitive FOUS had largely unchanged performance, with a SNR of 47.7 for straight fiber and a SNR of 36.8 at a bend radius of 3.0 mm. When integrated into the steerable guide catheter, the mean SNRs of the ultrasonic tracking signals recorded with the catheter in a cardiac phantom were similar for straight and fully deflected distal ends: 195 and 163. Conclusion: The FOUS fabricated from bend-insensitive fiber overcomes the bend restrictions associated with the FOUS fabricated from standard single mode fiber, thereby enabling its use in ultrasonic tracking in a wide range of cardiovascular devices.

Type: Article
Title: Bend-insensitive Fiber Optic Ultrasonic Tracking Probe for Cardiovascular Interventions
Open access status: An open access version is available from UCL Discovery
DOI: 10.1002/mp.16334
Publisher version: https://doi.org/10.1002/mp.16334
Language: English
Additional information: Copyright © 2023 The Authors. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
UCL classification: UCL
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng
URI: https://discovery-pp.ucl.ac.uk/id/eprint/10165319
Downloads since deposit
2,432Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item