The Impact of Electrogram Type and Conduction Velocity Estimation Techniques on Assessments of Conduction Velocity During Ventricular Substrate Mapping

Ehnesh, M; Tonko, J; Zolotarev, AM; Vigmond, EJ; Lambiase, PD; Roney, C; (2023) The Impact of Electrogram Type and Conduction Velocity Estimation Techniques on Assessments of Conduction Velocity During Ventricular Substrate Mapping. In: Computing in Cardiology 2023. Computing in Cardiology Green open access

[thumbnail of CinC2023-402.pdf]
Preview
Text
CinC2023-402.pdf - Published Version

Download (339kB) | Preview

Abstract

Conduction velocity (CV) mapping has the potential to identify critical regions of the ventricular substrate to target by ablation. However, it is unknown how electrogram modality and methodology affects CV estimauon. We aimed to test these effects in a control population for patients with structurally normal hearts and idiopathic ventricular ectopy. Electroanatomical contact mapping was performed using EnSite™ X (Abbott) with the Advisor HD Grid for patients with structurally normal hearts undergoing elective ventricular ectopy ablation. CV was assessed using four methods: 1. omnipolar wavespeed (Abbott); 2. local gradient estimation from the interpolated LAT field; 3. fitting a planar wavefront to LAT measurements; 4. fitting a circular wavefront and estimating CV. CV estimates were compared between methods. We analysed a total of 25 maps for 5 cases. Our results reveal than mean CV depends on CV estimation technique: e.g. for omnipolar recordings, wavespeed, Mean ± SD (m/s); 1.06 pm 3.36; gradient; 0.62 pm 0.40; fitting a planar wavefront; 0.94 pm 0.42; fitting a circular wavefront; 0.93 pm 0.90. Median pointwise difference is also large between EGMs and across techniques. Thus, in structurally normal hearts, CV estimation technique has a large effect on CV maps calculated during ventricular substrate mapping.

Type: Proceedings paper
Title: The Impact of Electrogram Type and Conduction Velocity Estimation Techniques on Assessments of Conduction Velocity During Ventricular Substrate Mapping
Event: 2023 Computing in Cardiology Conference
ISBN-13: 9798350382525
Open access status: An open access version is available from UCL Discovery
DOI: 10.22489/CinC.2023.402
Publisher version: http://dx.doi.org/10.22489/cinc.2023.402
Language: English
Additional information: CinC has been an open-access publication, in which copyright in each article is held by its authors, who grant permission to copy and redistribute their work with attribution, under the terms of the Creative Commons Attribution License. https://creativecommons.org/licenses/by/4.0/
UCL classification: UCL
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science
UCL > Provost and Vice Provost Offices > School of Life and Medical Sciences > Faculty of Population Health Sciences > Institute of Cardiovascular Science > Clinical Science
URI: https://discovery-pp.ucl.ac.uk/id/eprint/10186403
Downloads since deposit
1,140Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item