A highly miniaturized electron and ion energy spectrometer prototype for the rapid analysis of space plasmas.

Bedington, R; Kataria, DO; Smith, A; (2014) A highly miniaturized electron and ion energy spectrometer prototype for the rapid analysis of space plasmas. Review of Scientific Instruments , 85 (2) , Article 023305. 10.1063/1.4865842. Green open access

[thumbnail of CATS paper 1 1 unformatted.pdf] PDF
CATS paper 1 1 unformatted.pdf
Available under License : See the attached licence file.

Download (846kB)

Abstract

MEMS (Micro Electro-Mechanical Systems) plasma analyzers are a promising possibility for future space missions but conventional instrument designs are not necessarily well suited to micro-fabrication. Here, a candidate design for a MEMS-based instrument has been prototyped using electron-discharge machining. The device features 10 electrostatic analyzers that, with a single voltage applied to it, allow five different energies of electron and five different energies of positive ion to be simultaneously sampled. It has been simulated using SIMION and the electron response characteristics tested in an instrument calibration chamber. Small deviations found in the electrode spacing of the as-built prototype were found to have some effect on the electron response characteristics but do not significantly impede its performance.

Type: Article
Title: A highly miniaturized electron and ion energy spectrometer prototype for the rapid analysis of space plasmas.
Location: United States
Open access status: An open access version is available from UCL Discovery
DOI: 10.1063/1.4865842
Publisher version: http://dx.doi.org/10.1063/1.4865842
Language: English
Additional information: Copyright (2014) American Institute of Physics. This article may be downloaded for personal use only. Any other use requires prior permission of the author and the American Institute of Physics. The following article 'Bedington, R; Kataria, DO; Smith, A; (2014) A highly miniaturized electron and ion energy spectrometer prototype for the rapid analysis of space plasmas. Review of Scientific Instruments , 85 (2) , Article 023305.' and may be found at http://dx.doi.org/10.1063/1.4865842
UCL classification: UCL
UCL > Provost and Vice Provost Offices > UCL BEAMS
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Maths and Physical Sciences
UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Maths and Physical Sciences > Dept of Space and Climate Physics
URI: https://discovery-pp.ucl.ac.uk/id/eprint/1463752
Downloads since deposit
15,808Downloads
Download activity - last month
Download activity - last 12 months
Downloads by country - last 12 months

Archive Staff Only

View Item View Item