Song, Y;
Totz, J;
Thompson, S;
Johnsen, S;
Barratt, D;
Schneider, C;
Gurusamy, K;
... Clarkson, MJ; + view all
(2015)
Locally rigid, vessel-based registration for laparoscopic liver surgery.
International Journal of Computer Assisted Radiology and Surgery
, 10
(12)
pp. 1951-1961.
10.1007/s11548-015-1236-8.
Preview |
Text
art%3A10.1007%2Fs11548-015-1236-8.pdf Download (1MB) | Preview |
Abstract
Purpose: Laparoscopic liver resection has significant advantages over open surgery due to less patient trauma and faster recovery times, yet is difficult for most lesions due to the restricted field of view and lack of haptic feedback. Image guidance provides a potential solution but is challenging in a soft deforming organ such as the liver. In this paper, we therefore propose a laparoscopic ultrasound (LUS) image guidance system and study the feasibility of a locally rigid registration for laparoscopic liver surgery. Methods: We developed a real-time segmentation method to extract vessel centre points from calibrated, freehand, electromagnetically tracked, 2D LUS images. Using landmark-based initial registration and an optional iterative closest point (ICP) point-to-line registration, a vessel centre-line model extracted from preoperative computed tomography (CT) is registered to the ultrasound data during surgery. Results: Using the locally rigid ICP method, the RMS residual error when registering to a phantom was 0.7 mm, and the mean target registration error (TRE) for two in vivo porcine studies was 3.58 and 2.99 mm, respectively. Using the locally rigid landmark-based registration method gave a mean TRE of 4.23 mm using vessel centre lines derived from CT scans taken with pneumoperitoneum and 6.57 mm without pneumoperitoneum. Conclusion: In this paper we propose a practical image-guided surgery system based on locally rigid registration of a CT-derived model to vascular structures located with LUS. In a physical phantom and during porcine laparoscopic liver resection, we demonstrate accuracy of target location commensurate with surgical requirements. We conclude that locally rigid registration could be sufficient for practically useful image guidance in the near future.
Type: | Article |
---|---|
Title: | Locally rigid, vessel-based registration for laparoscopic liver surgery |
Open access status: | An open access version is available from UCL Discovery |
DOI: | 10.1007/s11548-015-1236-8 |
Publisher version: | http://dx.doi.org/10.1007/s11548-015-1236-8 |
Language: | English |
Additional information: | © The Author(s) 2015. This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. |
Keywords: | Registration; Ultrasound; Laparoscopic; Liver surgery |
UCL classification: | UCL UCL > Provost and Vice Provost Offices UCL > Provost and Vice Provost Offices > UCL BEAMS UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science UCL > Provost and Vice Provost Offices > UCL BEAMS > Faculty of Engineering Science > Dept of Med Phys and Biomedical Eng |
URI: | https://discovery-pp.ucl.ac.uk/id/eprint/1470775 |
Archive Staff Only